ADS-6000B
Technical parameters:
1,The blood pressure of the extremities was measured synchronously by the oscillometric method.
2,Measurement site: limbs: upper arm, lower limb ankle.
3,Arm circumference measurement range: 20-38cm.
4,The width range of lower limb measurement: 16cm-33cm.
5,Rated power: 115VA voltage 220V.
6,Performance requirements of blood pressure measurement:
Measurement range: 0kPa ~ 34.7kPa(0mmHg ~ 260mmHg);
Measurement accuracy: ±0.4kPa(±3mmHg);
Repeatability: measured under static continuous low pressure, the difference between the repeated measured readings at each point within the scale range is not greater than 0.533 kPa(4mmHg).
7,12.1 inch HD LCD screen, branded industrial control computer system.
8,Measurement parameters: baPWV (brachial-ankle pulse wave velocity), ABI (ankle brachial index), BAI (brachial-ankle index), HR(heart rhythm), PVR (pulse volume recording), systolic blood pressure SBP(limbs), diastolic blood pressure DBP(limbs), mean blood pressure MAP(limbs), pulse pressure PP(limbs).
9,Portable structural design, easy to move.
10,Equipped with lithium battery, capacity of 5200mAH.
11,Lithium battery working time up to 3 hours.
ADS-6000B Technical parameters of Arteriosclerosis Detector
ADS-6000B Arteriosclerosis Detector is an arterial stiffness and obstruction degree detector with complete independent intellectual property rights. The instrument uses synchronous compression to measure the blood pressure of the limbs noninvasively, simultaneously records and analyzes the pulse wave shape of the limbs, uses waveform analysis technology to extract the characteristic information that characterizes the blood vessel elasticity, calculates a number of indexes including ankle brachial index (ABI), brachial ankle artery pulse wave velocity (baPWV), and performs automatic analysis and comprehensive evaluation through the database data, to evaluate the health status of cardiovascular system and the degree of lower extremity artery obstruction, and provide comprehensive objective basis for early diagnosis of arteriosclerosis and early screening of cardiovascular and cerebrovascular diseases.